Fresenius Kabi Infusia SP Syringe Pump
Hardware Specifications
- Mechanism: Piston driver mechanism
- Accuracy: ±3%, ±1% device accuracy (without disposable variability)
- A/C Power: Yes
- D/C Power: Yes
- Battery: Lithium-ion polymer battery (11.1 Vdc, 2000 mAh)
- Battery Charger: Internal
- Battery Life: Approx. 8 hrs at 5 mL/hr
- Recharge Time: Pump OFF: <10 hrs, Pump ON: <14 hrs
- Weight: 5.1 lbs (2.3 kg)
- Dimensions: 9.1 x 6.1 x 5 in. (W x D x H)
- Syringe Sizes: 5, 10, 20, 30, 50/60 mL
- Display: Color LCD screen
- Screen Size: 4.1 x 1.3 in.
- Key Pad: Increment and decrement arrows
- Infusion Log: Biomed accessible, stores 5000 recent infusion and alarm events
- Calibration: Direct access from front panel
Performance Specifications
- Priming: Manual or via purge/bolus function
- KVO Rate: 1.0 mL/h or last delivery rate, whichever is lower
- Rate Titration: Yes (in rate mode)
- Loading Dose: Yes
- Delivery Modes: Rate Mode, Time Mode, Dose Mode (mcg/kg/min, mg/kg/h)
- Drug Mass: 0.1 to 999.9 mg (0.1 mg increments)
- Dose:
- 0.01 to 99.9 mcg/kg/min
- 0.01 to 99.9 mg/kg/h
- Patient Weight:
- 0.0 to 99.9 kg (0.1 kg increments)
- 100 to 300 kg (1 kg increments)
- Solution Volume: 0.1 to 1999.9 mL (0.1 mL increments)
Safety Systems
- Alarm Volume: Adjustable – Low, Medium, High
- Occlusion Settings:
- High: 900 - 1125 mmHg
- Medium: 600 - 900 mmHg
- Low: 375 - 600 mmHg
- Post-Occlusion Bolus Reduction: Yes
- Pump Status Indicators:
- Green light for infusion in progress
- Red alarm alert for warnings
- KVO and Bolus indicator
- Visual, text, and sound alarm notifications
- Syringe Position Control: Clasp check, flange insertion, plunger head detection
- Infusion Control Alarms: Occlusion pressure, end of infusion, near empty
- Device Control Alarms: Disengaged syringe, low battery, technical malfunction
- Air Sensor: Not applicable
- Key Pad Lock: Automatic
Technical Specifications
- Compliance Standards: EN/IEC 60601-1, 60601-2-24, 60601-1-2, IEC 60529:2001
- Waterproofness: IPX4
- Electrical Compliance:
- Leakage current protection: CF type
- Electric shock protection: Class I
- Power Supply: 100 – 240 Vac, 50/60 Hz, 35 VA
- Environmental: RoHS Compliant (2017)
- EMC: NRTL Mark (UL), EN/IEC 60601-1-2/AC2010, EN/IEC 60601-2-24